Changzhou Tonghui Electronic Co., Ltd.
Products
Contact Us
  • Contact Person : Mr. Lu David
  • Company Name : Changzhou Tonghui Electronic Co., Ltd.
  • Tel : 86-519-85195566
  • Fax : 86-519-85109972
  • Address : Jiangsu,Changzhou,No.3 Tianshan Road,New District,Changzhou,Jiangsu
  • Country/Region : China
  • Zip : 213022

Automatic Transformer Test System TH2819XA

Automatic Transformer Test System TH2819XA
Product Detailed
Related Categories:Testing Equipment
1. 20-channel transformer scanning measurement 2. Frequency up to 200kHz 3. 10-point list sweep 4. Operational DC Bias sourc

Introduction

TH2819XA is an automatic transformer multi-parameter measurement instrument used for quality control of electronic transformer over a wide range of frequency (20 Hz to 200 kHz). Almost all low-voltage transformer parameters can be scanned, including the primary parameters of inductance L, leakage inductance Lk, turns-ratio Tr, DC resistance DCR, impedance Z, and capacitance (between windings) C etc. and the secondary parameters of quality Factor Q , ESR and phase etc.,  pin short can also be detected as well.

Model

TH2819XA

System

20-channel transformer scanning measurement + LCR meter

Measurement

Parameters

LCR

|Z|, |Y|, C, L, X, B, R, G, D, Q, θ DCR, turns ratio, phase, leakage inductance

Transformer

turns ratio, turns, phase, L, C, Lk, Q, ACR, DCR, balance, pin short

Basic accuracy

LCRZ, DCR: 0.1%

Turns ratio: 0.5%

Equivalent

Serial, Parallel

Math function

Absolute value deviation, percentage deviation

Range

Auto, Hold, Manual

Trigger

Internal, Manual, External

Measurement time

Fast,32ms, Medium, 90ms, Slow, 650ms

Average

1—255

Delay

0—60s, with the step of 1ms

Correction

Open/Short point-frequency, full-frequency , Load

Measure terminal

5-terminal

Model

Direct, Δ , Δ%, V/I (voltage/current monitor)

Display

320×240 dot matrix graphic LCD display with 6-digit resolution

Test signal

Test signal frequency

20 Hz-200 kHz, resolution 10mHz

Output impedance

Turns measurement

Test level ≤ 2V, 10Ω, 30 Ω, 100 Ω

> 2V, 30 Ω, 100 Ω

Others

30 Ω, 100 Ω

Test signal level

LCR:  5 mV-2 V  Accuracy: 10%

Turns: 5 mV-4 V   Accuracy: 10%

Auto level range: 10 mV-1 V Accuracy: 5%

Level min. resolution

1mV

DC bias source

Internal

0V, 1.5V, 2V, accuracy 1%

Optional

TH1902A 0-10V DC bias voltage, 0-100mA DC bias current

TH1902B 0-1A DC bias current

Measurement display range

|Z|, R, X

0.00001 Ω — 99.9999 MΩ

DCR

0.001 Ω — 99.9999 MΩ

|Y|, G, B

0.00001µS — 99.9999 S

C

0.00001 pF — 9.9999 F

L

0.00001 µH — 9999.99 H

D

0.00001 — 9.9999

Q

0.00001 — 9999.9

θ (DEG)

-179.999º — 179.999 º

θ (RAD)

-3.14159 — 3.14159

Turn-Ratio

1:0.01—100:1

Δ%

-999.999% — 999.999%

Others

Comparator

10-bin sorting, bin counter

Storage

20 sets of LCRZ settings

40 sets of scanning settings

Interface

RS232C, USB HOST, Scanner

( function(){ var aBigImageSrc = [ 'http://i01.i.aliimg.com/photo/v2/395345487_1/Automatic_Transformer_Test_System_TH2819XA.jpg' , 'http://i01.i.aliimg.com/photo/v2/395345487_2/Automatic_Transformer_Test_System_TH2819XA.jpg' , 'http://i01.i.aliimg.com/photo/v2/395345487_3/Automatic_Transformer_Test_System_TH2819XA.jpg' ]; var oProductShow = new AE.run.minisite.productShow(); oProductShow.init( { 'aImgSrc': aBigImageSrc } ); } )(); Automatic Transformer Test System TH2819XA